Echotel model 355 two-wire loop-powered ultrasonic non-contact transmitter

Downloads

Echotel® model 355 two-wire loop-powered ultrasonic non-contact transmitter

 

Echotel® Model 355 is a loop-powered, non-contact-type ultrasonic transmitter that performs liquid level, volume or open channel flow measurement. HART® communications are offered with a PACTware™ PC program for remote configuration and troubleshooting.

Ultrasonic non-contact

Technology

The level measurement is made by emitting an ultrasonic pulse from the transducer, then measuring the time required for the echo to reflect from the liquid surface and return to the transducer. The powerful electronics measure the time of the round trip pulse and, by knowing the speed of sound, calculate the distance. Since speed of sound is temperature dependent, the transducer also measures the temperature in the vessel to provide compensation for changing temperature.

  • PACTware PC program using HART Communications for remote configuration, advanced diagnostics, performance trending and troubleshooting
  • Menu-driven 4-pushbutton, 2-line x 16 character LCD user interface
  • False target rejection for identification of true echo from liquid surface
  • Common tank shapes and 20-point custom table for volume measurement
  • Resettable and non-resettable 7-digit flow totalizers
  • Flume and weir curves for open channel flow measurement, in addition to a 20-point custom table and a generic discharge flow equation
Chemical
  • Lime slurry level
  • Neutralization
Common applications
  • Flow: Open channel flow
Crude oil
  • Drilling fluid tanks
  • Process & Field storage tanks
Nuclear power
  • Cooling tower intake & basin levels
Power generation
  • Condensate storage
  • Cooling tower intake & basin levels
  • Fuel oil storage
  • Water services
  • Water wash tanks
Pulp & paper
  • Digester blow tanks
  • Mill water storage
  • Source water & bar screens
Renewable energy
  • Biofuel
  • Biogas
  • Geothermal
Water & wastewater
  • Carbon slurry level
  • Clarifier level
  • Digester level
  • Filter tank level
  • Influent/effluent flow
  • Lift station pump control
  • Lime slurry level
  • Sludge level
  • Splitter box level
  • Water services
  • Water storage tanks
  • Water wash tanks
ATEX Ex d, Ex ia
CCOE Ex ia
cFMus IS, NI
Inmetro Ex d, Ex ia
SIL SIL 1 (1oo1)

Other approvals are available.

Ultrasonic pulses are emitted from the transducer, and the time required for the echo to reflect from the liquid surface and return to the transducer is measured.

+ Features
  • PACTware PC program using HART Communications for remote configuration, advanced diagnostics, performance trending and troubleshooting
  • Menu-driven 4-pushbutton, 2-line x 16 character LCD user interface
  • False target rejection for identification of true echo from liquid surface
  • Common tank shapes and 20-point custom table for volume measurement
  • Resettable and non-resettable 7-digit flow totalizers
  • Flume and weir curves for open channel flow measurement, in addition to a 20-point custom table and a generic discharge flow equation
+ Applications
Chemical
  • Lime slurry level
  • Neutralization
Common applications
  • Flow: Open channel flow
Crude oil
  • Drilling fluid tanks
  • Process & Field storage tanks
Nuclear power
  • Cooling tower intake & basin levels
Power generation
  • Condensate storage
  • Cooling tower intake & basin levels
  • Fuel oil storage
  • Water services
  • Water wash tanks
Pulp & paper
  • Digester blow tanks
  • Mill water storage
  • Source water & bar screens
Renewable energy
  • Biofuel
  • Biogas
  • Geothermal
Water & wastewater
  • Carbon slurry level
  • Clarifier level
  • Digester level
  • Filter tank level
  • Influent/effluent flow
  • Lift station pump control
  • Lime slurry level
  • Sludge level
  • Splitter box level
  • Water services
  • Water storage tanks
  • Water wash tanks
+ Approvals
ATEX Ex d, Ex ia
CCOE Ex ia
cFMus IS, NI
Inmetro Ex d, Ex ia
SIL SIL 1 (1oo1)

Other approvals are available.

+ Operating principle

Ultrasonic pulses are emitted from the transducer, and the time required for the echo to reflect from the liquid surface and return to the transducer is measured.

JOLAC ENGINEERING SDN BHD (“JOLAC”) was incorporated as private limited company in Kuala Lumpur, Malaysia under the Companies Act 1965 on 16th December 1982.